The FilmSense FS-1 integration watches a Film Sense ellipsometer over its network command interface (FS-API) and streams each dynamic measurement run to Atomscale as an ellipsometry data item. It only observes: you start and stop measurements in the Film Sense software, and the integration never changes the instrument’s model or acquisition settings. Stopping the integration doesn’t stop acquisition on the instrument. For the steps every integration shares, see Working with Integrations.

How a Run Is Captured

  • Run start: Every Polling Interval, the integration asks the FS-1 whether dynamic measurements are being acquired. When acquisition starts, a new run begins.
  • Measurements: Each dynamic measurement point, with its fitted parameters and the instrument’s Time, becomes one row. Rows are uploaded in chunks of Points per Chunk, released every Delivery Interval.
  • Run end: The run ends once acquisition has stopped and no new points have arrived for Idle Timeout, so trailing measurements are still collected.
  • In Atomscale: Each run becomes one ellipsometry data item named filmsense_ followed by the run’s start time in Unix milliseconds.
With Raw Dynamic Export on, the integration also downloads the instrument’s native dynamic data file during the run (at most once per Raw Export Interval) and again when the run ends. With Upload Final Native Copy on, the final file is stored as a separate ellipsometry item linked to the streamed run.

Before You Start

  • The desktop app on Windows 10 or later. The Atomscale CLI doesn’t run this integration.
  • Film Sense software version 3.77 or higher on the FS-1, which the FS-API requires.
  • The gateway on the same local network as the FS-1, either with an Ethernet cable directly to the ellipsometer or through a network switch or router. By default the FS-1 answers at 169.254.1.1 (or fs.local) on TCP port 4000.
  • If you changed the FS-1’s hostname or set a static IP address in the Film Sense software, use that address instead.

Set Up the Integration

To have the Gateway assistant install and configure FilmSense FS-1, give it the instrument’s address and port. Use Test connection in the configuration form to verify the connection.
1

Add the integration

Open Integrations, click Add Integration, and click Add on FilmSense FS-1. Add one copy per instrument.
2

Enter the instrument address

Set Instrument Host and Instrument Port. The defaults, 169.254.1.1 and 4000, match an FS-1 with its factory network settings.
3

Test the connection

Click Test connection in the Instrument Host field. It sends the FS-API communication check, which doesn’t change anything on the instrument. If it fails, the reason appears under the field.
4

Apply and start

Click Apply Configuration, then Start Integration.
5

Verify with a run

Start dynamic measurements in the Film Sense software. The integration’s Activity Logs show FilmSense run started, and FilmSense run ended after acquisition stops and the idle timeout passes.
To try the integration without creating new uploads, turn on Dry Run. Runs are captured on the gateway only. Data already queued for upload is still delivered.

Settings Reference

All interval and timeout fields accept 0.001 to 86,400 seconds.

Troubleshooting

Check the integration’s Activity Logs on the gateway.