In either mode, a change in image size closes the current data item and opens a new one.
Frame Activity
A frame counts as active when its average brightness is at least Minimum Mean Intensity. Dark frames neither open a data item nor keep it open, but dark frames inside an open data item are still kept. This keeps a camera that streams between growths, for example with the beam off, from recording empty data items. Leave Minimum Mean Intensity blank and the integration learns it. Each time capture starts, the threshold begins at 0, so every frame counts. Once the last half hour has shown the beam both off and on, the integration sets the threshold just above the darkest frames. It lowers the threshold if darker frames arrive, but never raises it while capturing, so dim patterns are kept. It starts over after the camera sends nothing for half an hour. To set it yourself, enter a value from 0 to 255; 0 counts every frame as active. Use Auto Manage to hand it back to the integration. Set Session Gap longer than normal pauses within a run and shorter than the gap between runs.Following kSA 400
1
Enable kSAComm in kSA 400
In kSA 400, open Options > General > Chamber Interface and select kSAComm (default port
1800). Make sure Save settings at program exit is enabled, then restart kSA 400. It accepts
connections only after the restart.
2
Switch the boundary mode
Stop the integration, open Configure, and set Data Item Boundary Mode to kSA
Acquisition Status. If kSA 400 runs on a different computer, set kSA Communication Host to
its address.
3
Apply and start
Apply the configuration and start the integration. A data item opens when kSA 400 starts
acquiring and closes when acquisition stops.
What Gets Uploaded
- 8-bit grayscale. Frames are converted before upload. Color (Bayer) frames are averaged to grayscale at half width and height. For high bit-depth cameras whose images look too dark, set White Point (advanced) to the pixel value that should map to white.
- Frame rate. Up to 120 frames per second. The rate is measured once at the start of each data item. Upload Frame Rate Limit caps how many frames per second upload; extra frames are skipped, and the camera’s own rate doesn’t change.
- Spatial Downsampling. Halves width and height before upload, either by dropping pixels or averaging each 2 × 2 block (Downsampling Method). Brightness checks still use the full image.
- Stage rotation. With Stage Rotation set, every data item is analyzed as rotating RHEED at that speed. Left blank, Atomscale detects whether the stage rotates.
- Camera settings. Each data item records the camera’s exposure, gain, pixel format, image size, frame rate, model, and serial when it opens.